Various areas of application place different demands on the technology. The CAP201 CLINIC is designed for clinical applications with a high measurement accuracy. Whether it’s measurement accuracy, resistance to environmental conditions, weight or flexibility in assembly - upon request, we can customize the capnography module to make it easier for you to integrate it into your technology. We offer optimized versions for all common areas of application.
Specification | Values |
Measurement range: | 0 vol% to 16 vol% |
Measurement accuracy: | 0 vol% to 3.7 vol%: ± 0.26 vol% > 3.7 vol%: ± 7 % of real concentration |
CO2 value output rate: | 8 Hz or 40 Hz (selectable) |
Rise time T10-90: | < 150 ms @ 80 ml / min |
Start-up time (warm-up): | < 5 s |
Power consumption | ~ 1.0 W (+ 0.3 W with pump) |
Size of board: | 70 mm x 40 mm x 26 mm (LxWxH) |
Weight: | 69 g (incl. pump) |
High accuracy sidestream CO₂ measurement
Low energy consumption and small footprint
Calculation of fiCO₂, etCO₂ and breathing rate
Compensation of aging and environmental effects
Occlusion and breathing detection
No recalibration in field required